MSACL 

65. Increased Process Efficiency in UPLC-MS/MS Data Review Through Use of Rules-Based Software Automation
Poster: Mon 2:00-3:00PM
Joseph Ervin
AIT Laboratories
View Long Abstract
A previously manual data review process for opiates in urine using UPLC-MS/MS was automated via Indigo BioSystems' ASCENT software system. Ascent applied data quality rules and determined that approximately 90% of results adhered to quality standards and were releasable without further human review. Additionally all problematic results needing human intervention were flagged by one of the rule scripts due to data unacceptability. This process has further implications as result quality can be increased through consistent application of rules based on quality standards.
Email: Jervin@aitlabs.com