MSACL 2023 : Miller

MSACL 2023 Abstract

Self-Classified Topic Area(s): Assays Leveraging MS > Various OTHER

Hydrolysis Monitoring for Accurate Clinical Measurement of Total Carnitine

Artis Miller, Deborah Cooper, Katherine Blakeman, Kirsten Bush, Stephen R. Master
Michael Palmieri Laboratory for Metabolic and Advanced Diagnostics Department of Pathology and Laboratory MedicineChildren’s Hospital of Philadelphia, Philadelphia, PA, USA

 Artis Miller, BS Biology (Presenter)
The Children’s Hospital of Philadelphia

>> POSTER (PDF)

Relevant Financial Disclosures (within past 24 months)
No relevant financial relationship(s) to disclose.

Abstract

Introduction:
Carnitine plays a crucial role in transporting fatty acids into the mitochondrial matrix, and conjugation of free carnitine to an acyl group to form an acylcarnitine is a necessary step in this process. Free and total carnitine are routinely measured in order to diagnose carnitine deficiency as well as inherent mitochondrial defects. Free carnitine is measured by tandem mass spectrometry after a simple protein crash, and it represents the fraction of carnitine unconjugated to fatty acid. Similarly, total carnitine is measured by flow injection-MS/MS after alkaline hydrolysis to liberate carnitine from associated acyl groups. Acylcarnitine is calculated as the difference between total and free carnitine. Because of the dependence on efficient alkaline hydrolysis, both the total and acylcarnitine values may be artifactually low if carnitine is not fully liberated from bound acyl groups. Importantly, the ERNDIM quality assurance program only tests free carnitine, and thus it is possible to miss systematic problems in the hydrolysis step if total carnitine is also measured.

Methods:
Our objective was to develop an automated flag to detect samples in which hydrolysis was incomplete. Our underivatized free/total carnitine method monitors free carnitine (C0) by flow injection-MS/MS (quantifier: 162.25 -> 85.17; qualifier: 162.25 -> 103.19) and a d3-carnitine IS. We added an additional MRM transition to monitor acetylcarnitine (C2) (204.25 -> 85.17), which was also normalized using the C0 IS. We monitored the decrease in normalized peak area of C2 in free vs. total carnitine samples in order to detect appropriate hydrolysis. This calculation was also added to Indigo Ascent, and a rule was constructed to flag samples with insufficient % decreases in C2.

Results:
Although many acylcarnitines are present only in low abundance in healthy individuals, acytylcarnitine (C2) is normally present at levels of approximately 4-21 umol/L, with notably high values present with ketosis. As such, measuring its decrease as a result of alkaline hydrolysis may provide a convenient way to detect the efficiency of this step. We added an MRM transition for C2 to our flow injection free/total carnitine assay, and because of co-elution during flow injection we used the C0 IS to provide overall normalization of the C2 peak (designated nC2).

During assay validation, we identified 2 outlier measurements of a high control sample. Interestingly, these outliers showed a substantially higher (total nC2) / (free nC2) ratio (0.62-0.67) compared with other samples (<0.1), consistent with suboptimal hydrolysis. This corresponded with lower measured total carnitine in these results compared with the other values in the precision study, as would also be expected for incomplete hydrolysis. Free carnitine values were comparable across the precision replicates, showing that this did not reflect a global change in the ability to measure carnitine. These data provided evidence that total / free nC2 is a useful measure to identify samples in which incomplete hydrolysis has occurred. Subsequently, this measurement has been implemented as a rule in Indigo Ascent, and we show a clinical example where a sample would have been reported with an artifactually low total carnitine if the nC2 ratio had not flagged incomplete hydrolysis.

Conclusions:
Monitoring the %decrease of normalized C2 abundance between free and total carnitine samples using a rule-based flag in our MS processing software (Indigo Ascent) provides a convenient and automated method to ensure the integrity of the hydrolysis step in free/total carnitine assays.